武汉市所在地
气体种类 | 测量范围 | 示值允许误差 | |
误差 | 相对误差 | ||
HC | (0~2000)×10-6VOL | ±4×10-6VOL | ±3% |
(2001~5000)×10-6VOL | - | ±5% | |
(5001~9999)×10-6VOL | - | ±10% | |
CO | (0.00~10.00)×10-2VOL | ±0.02×10-2VOL | ±3% |
(10.01~14.00)×10-2VOL | - | ±5% | |
CO2 | (0.0~16.0)×10-2VOL | ±0.3×10-2VOL | ±3% |
(16.01~18.0)×10-2VOL | --- | ±5% | |
NO | (0~4000)×10-6VOL | ±25×10-4VOL | ±4% |
(4001~5000)×10-6VOL | --- | ±8% | |
NO2 | (0~1000)×10-6VOL | ±25×10-6VOL | ±4% |
O2 | (0.0~25.0)×10-2VOL | ±0.1×10-2VOL | ±5% |
注:表中所列误差和相对误差,满足其中一项要求即可 | |||
分辨率 | HC/NO/NO2:1ppm;CO2/CO/O2:0.01%; | ||
响应时间 | HC/CO/CO2:T90≤8s;NO/NO2:T90≤10s;O2:T90≤12s | ||
预热时间 | 30min | ||
通信 | RS-485/232,可连接ZK-8000系列尾气及烟度分析设备 | ||
电源 | AC100V~240V,50Hz±2% |
气体种类 | 测量范围 | 示值允许误差 | |
误差 | 相对误差 | ||
HC | (0~2000)×10-6VOL | ±4×10-6VOL | ±3% |
(2001~5000)×10-6VOL | - | ±5% | |
(5001~9999)×10-6VOL | - | ±10% | |
CO | (0.00~10.00)×10-2VOL | ±0.02×10-2VOL | ±3% |
(10.01~14.00)×10-2VOL | - | ±5% | |
CO2 | (0.0~16.0)×10-2VOL | ±0.3×10-2VOL | ±3% |
(16.01~18.0)×10-2VOL | --- | ±5% | |
NO | (0~4000)×10-6VOL | ±25×10-4VOL | ±4% |
(4001~5000)×10-6VOL | --- | ±8% | |
NO2 | (0~1000)×10-6VOL | ±25×10-6VOL | ±4% |
O2 | (0.0~25.0)×10-2VOL | ±0.1×10-2VOL | ±5% |
注:表中所列误差和相对误差,满足其中一项要求即可 | |||
分辨率 | HC/NO/NO2:1ppm;CO2/CO/O2:0.01%; | ||
响应时间 | HC/CO/CO2:T90≤8s;NO/NO2:T90≤10s;O2:T90≤12s | ||
预热时间 | 30min | ||
通信 | RS-485/232,可连接ZK-8000系列尾气及烟度分析设备 | ||
电源 | AC100V~240V,50Hz±2% |